next up previous contents
Next: IWC Up: RMC Spectrometer Previous: RMC Spectrometer   Contents

Lead Converter

A 1.1 mm thick, 35 cm long and 13 cm radius cylindrical lead sheet was wrapped around the $A, A'$ scintillators. This lead sheet converted the real photon-pairs into pairs of $e^+e^-$ tracks.


Table 3.2: RMC Spectrometer dimensions (26). CW denotes the cell half-width at the radial center of the superlayer and S denotes the spacing between IWC anode wires or the cathode strips.
       
Component Sectors Inner Radius (cm) Dimension
    $r$ $\ell \times 2\pi r \times t$
       
A Scint. 4 arcs 12.20 508$\times$191.6$\times$3.2 mm
A' Scint. 4 arcs 12.60 508$\times$197.9$\times$3.2 mm
B Scint. 12 sectors 15.20 483$\times$ 81.5$\times$3.2 mm
C Scint. 12 sectors 15.80 457$\times$ 84.7$\times$3.2 mm
D Scint. 16 sectors 59.37 940$\times$236.0$\times$6.4 mm
       
Pb Converter 1 sheet 13.00 817$\times$ 351$\times$1 mm
       
IWC      
Inner Cathode 384 strips 26.47 -45$^\circ$ S = 3.06 mm
Anode 768 wires 27.02 Axial S = 2.21 mm
Outer Cathode 384 strips 27.57 +45$^\circ$ S = 3.19 mm
       
Drift Chamber      
Inner carbon      
fiber cylinder   30.20 90$\times$0.075 cm
Superlayer 1 56 cells 34.70 CW = 1.947 cm
Superlayer 2 64 cells 40.80 CW = 2.003 cm
Superlayer 3      
(stereo) 72 cells 46.90 CW = 2.046 cm
Superlayer 4 80 cells 53.00 CW = 2.081 cm
Outer carbon      
fiber cylinder   58.00 90$\times$0.1 cm
       


The optimum thickness of the lead sheet was determined from Monte Carlo simulations (30) of the photon energy resolution ($\Delta E/E$). The energy resolution decreases with increasing thickness while the acceptance of photon detection ( $\epsilon \Omega$) increases with the thickness. The optimum thickness was found to be 1.1 mm from RMC GEANT Monte Carlo simulations. Thus a 1.1 mm lead sheet was used as the photon-converter. The actual thickness of the lead sheet used was also measured. 15 sample measurements with a digital caliper (small scale division=0.0001 in) yielded a mean thickness of 1.120 mm ( $\sigma_{I}$ = 0.025 mm) with standard error of measurement $\sigma$=$\sigma_I$/$\sqrt{14}$ = 0.007 mm, t = 1.120 $\pm$0.007 mm. When this thickness was calculated from the measured weight ($m$ = 1175.4 $\pm$ 14.2 $g$) of a sample (40 $\times$ 23 cm$^2$) of the lead sheet used for the lead converter;

\begin{eqnarray*}
m = \rho \times Vol. = 11.35 g/{\rm cm}^3
\times 40 {\rm cm} \times 23 {\rm cm}
\times t \quad ,
\end{eqnarray*}



the thickness was found to be $t$ = 1.126 $\pm$ 0.015 mm, where the dynamic propagation error is obtained from:

\begin{eqnarray*}
\frac{\delta t}{t} = \sqrt{ (\frac{\delta W}{W})^2 +(\frac{\de...
...175.8})^2 +(\frac{0.1}{40.0})^2
+(\frac{0.1}{23.0})^2 } = 0.013
\end{eqnarray*}



The $e^+e^-$ pairs converted from the photons followed curved tracks due to the axial magnetic field and were tracked by the multi-wire proportional chamber (IWC) and the large volume cylindrical drift chamber (DC).


next up previous contents
Next: IWC Up: RMC Spectrometer Previous: RMC Spectrometer   Contents
Sugata Tripathi 2004-03-27